How to deal with the problem that the swing pipe of the rain test machine cannot return to its original position

2025-05-16

Latest company news about How to deal with the problem that the swing pipe of the rain test machine cannot return to its original position

The rain test machine is mainly composed of main cabinet, auxiliary cabinet, swing pipe, stepper motor, water pump, control system and other components, among which the swing pipe is the key factor to determine the model of the rain test machine.


According to the standard of 200mm spacing between the swing pipe of the rain test machine and the sample shell, the appropriate swing pipe can be selected from R200, R400, R600, R800, R1000, R1200, R1400, and R1600.


Then the site where the rain test machine is placed should also be waterproofed, such as waterproof coating on the ground and walls, floor drains and drainage channels on the ground, and waterproof series electrical appliances in the room.


After using the rain test machine for a period of time, some customers found that after clicking the swing pipe to zero in the control system, there was still a certain error with the specified scale after the swing pipe was zeroed. How to calibrate in this case?


Mr. Wang from Yuexin After-sales Service Department is here to answer your questions:


1. Reason: After the rain test machine has been used for a period of time, it has not been adjusted and corrected in time due to excessive frequency of use or collision during use, resulting in deviation in the position of the swing tube and failure to reach the zero position.


2. Solution:

  • Check whether the positions of the main cabinet and the auxiliary cabinet are at the same horizontal position. If not, please adjust them in time.
  • Check whether the spacing between the main cabinet and the auxiliary cabinet causes the deformation of the swing tube. If deformation is found, adjust the spacing in time.
  • Check whether the casters of the main cabinet and the auxiliary cabinet are fixed. To fix them, please lock the four casters in time.
  • Check the error value between the swing tube and the zero angle, open the main cabinet, manually adjust the stepper motor induction electromagnetic sheet, and fine-tune the angle until the swing tube is perpendicular to the sample turntable and overlaps with the zero position.
  • Repeatedly test whether the swing tube can return to zero after swinging up and down. If it is not zero, continue to debug according to the previous step until the swing tube returns to zero.


The above is the solution to the problem that the oscillating pipe of the rain test machine does not return to zero. I hope it will be helpful to everyone. If you encounter other reasons, you can contact our online after-sales service department at any time to serve you.


Yuexin-focuses on the research and development and production of rain test machines, and looks forward to every communication with you.


This article is an original article of Guangzhou Yuexin Testing Equipment Co., Ltd. and may not be reproduced without permission.